D234VRC Secura VR
Implantable Cardioverter Defibrillator (ICD)
D234VRC Secura VR
Information

Device Survival Probability

SECURA_VR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 162.0 mo              
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.9 99.8 99.8 99.7 99.6 99.6 99.6 99.6 99.6 99.6 99.6              
% Including Normal Battery Depletion 99.8 99.6 99.3 99.0 98.6 98.0 96.2 91.1 77.1 54.5 46.1 44.3 42.9 42.6              
#   17640 16331 15178 14073 12961 11846 10616 8586 5565 2829 1791 1271 519 108              
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35 J
More
Connector Style Cx Style
Serial Number Prefix PZF
Mass 68 g
Volume 37 cc
Waveform Biphasic
X-Ray ID PUG
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.3 4.7 4.9 5.0
Quarterly 6.7 7.5 8.1 8.4
Semi-annual 7.8 8.9 9.8 10.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release
CE Approval Date2008-03-14
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release
CE Approval Date2008-03-14
Estimated WW Distribution12761
Normal Battery Depletions23
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 34
Battery 8
Electrical Component 18
Possible Early Battery Depletion 3
Software/Firmware 5
Therapy Function Compromised 10
Battery 5
Electrical Component 5
Information

EOS Indication

The Prolonged Service Period (PSP) is the time between the RRT and EOS. The PSP is defined as 3 months assuming the following conditions: 100% VVI pacing at 60 min-1, 2.5 V RV pacing amplitude; 0.4 ms pulse width; 600 Ω pacing load; and 6 full-energy charges. The EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurements.

Information

Charge Time

D204VRM_D214VRM_D224VRC_D234VRC_CHRG
Loading...

Data as of December 1, 2023

www.medtronic.com/productperformance 951051