D234TRK Consulta CRT-D
Cardiac Resynchronization Therapy - Defibrillation (CRT-Defib)
D234TRK Consulta CRT-D
Information

Device Survival Probability

CONSULTA_CRT_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 74.0 mo      
% Excluding Normal Battery Depletion 100.0 99.7 99.3 98.8 98.7 98.5 98.5      
% Including Normal Battery Depletion 99.6 98.3 92.8 80.7 60.4 31.9 18.6      
#   57528 51958 42528 30282 12359 883 113      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type CRT-Defib
NBD Code DDE-DDDR
Max. Delivered Energy 35 J
More
Connector Style CRT-D Style
Serial Number Prefix PYZ
Mass 68 g
Volume 38 cc
Waveform Biphasic
X-Ray ID PUG
Hide this content
CRT-D Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.2 3.8 4.4 4.7
Quarterly 4.4 5.5 6.8 7.5
Semi-annual 4.8 6.2 7.9 9.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Integrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release
CE Approval Date2008-03-14
Registered USA Implants2
Estimated Active USA Implants1
Normal Battery Depletions
US Market Release
CE Approval Date2008-03-14
Estimated WW Distribution24477
Normal Battery Depletions413
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 52
Electrical Component 12
Possible Early Battery Depletion 39
Software/Firmware 1
Therapy Function Compromised 5
Electrical Component 4
Software/Firmware 1
Information

Recommended Replacement (RRT)

Battery voltage less than 2.63 V or Charge Time greater than 16 seconds.

Information

Charge Time

D204TRM_D214TRM_D224TRK_D234TRK_CHRG
Loading...

Data as of November 2, 2015

www.medtronic.com/productperformance 376051