D224VRC Secura VR
Implantable Cardioverter Defibrillator (ICD)
D224VRC Secura VR
Information

Device Survival Probability

SECURA_VR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr    
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.9 99.8 99.8 99.8 99.7    
% Including Normal Battery Depletion 99.8 99.6 99.4 99.1 98.8 98.2 96.0 91.2    
#   18312 17108 16055 14130 11803 8411 3747 207    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35 J
More
Connector Style Cx Style
Serial Number Prefix PUX
Mass 68g
Volume 37cc
Waveform Biphasic
X-Ray ID PUG
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.3 4.7 4.9 5.0
Quarterly 6.7 7.5 8.1 8.4
Semi-annual 7.8 8.9 9.8 10.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

The Prolonged Service Period (PSP) is the time between the RRT and EOS. The PSP is defined as 3 months assuming the following conditions: 100% VVI pacing at 60 min-1, 2.5 V RV pacing amplitude; 0.4 ms pulse width; 600 Ω pacing load; and 6 full-energy charges. The EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Distribution Data

US Market Release2008-09-15
CE Approval Date
Registered USA Implants20060
Estimated Active USA Implants12870
Normal Battery Depletions230
US Market Release2008-09-15
CE Approval Date
Estimated WW Distribution20470
Normal Battery Depletions231
Information

Malfunctions (USA)

Therapy Function Not Compromised 31
Battery 12
Electrical Component 8
Possible Early Battery Depletion 8
Software/Firmware 2
Other 1
Therapy Function Compromised 6
Electrical Component 5
Software/Firmware 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 32
Battery 12
Electrical Component 8
Possible Early Battery Depletion 8
Software/Firmware 2
Other 2
Therapy Function Compromised 6
Electrical Component 5
Software/Firmware 1
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurements.

Information

Charge Time

D204VRM_D214VRM_D224VRC_D234VRC_CHRG
Loading...

Data as of February 3, 2017

www.medtronic.com/productperformance 476050