D224VRC Secura VR
Implantable Cardioverter Defibrillator (ICD)
D224VRC Secura VR
Information

Device Survival Probability


SECURA_VR_SURV
    1 yr 2 yr 3 yr 4 yr at 53 mo          
% Excluding Normal Battery Depletion 100.0 99.9 99.8 99.8 99.8          
% Including Normal Battery Depletion 99.9 99.6 99.3 99.3 99.3          
#   17236 13001 6685 1535 132          
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35J
More
Connector Style Cx Style
Serial Number Prefix PUX
Mass 68g
Volume 37cc
Waveform Biphasic
X-Ray ID PUG
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.3 4.7 4.9 5.0
Quarterly 6.7 7.5 8.0 8.4
Semi-annual 7.8 8.9 9.8 10.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

The Prolonged Service Period (PSP) is the time between the RRT and EOS. The PSP is defined as 3 months assuming the following conditions: 100% VVI pacing at 60 min-1, 2.5 V RV pacing amplitude; 0.4 ms pulse width; 600 Ω pacing load; and 6 full-energy charges. The EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Distribution Data

US Market Release2008-09-15
CE Approval Date
Registered USA Implants19317
Estimated Active USA Implants16326
Normal Battery Depletions27
US Market Release2008-09-15
CE Approval Date
Estimated WW Distribution19880
Normal Battery Depletions28
Information

Malfunctions (USA)

Therapy Function Not Compromised 17
Electrical Component 3
Possible Early Battery Depletion 5
Software/Firmware 2
Other 7
Therapy Function Compromised 4
Electrical Component 3
Software/Firmware 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 18
Electrical Component 3
Possible Early Battery Depletion 5
Software/Firmware 2
Other 8
Therapy Function Compromised 4
Electrical Component 3
Software/Firmware 1
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurements.

Information

Charge Time


D204VRM_D214VRM_D224VRC_D234VRC_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050