D204TRM Consulta CRT-D
Cardiac Resynchronization Therapy - Defibrillation (CRT-Defib)
D204TRM Consulta CRT-D
Information

Device Survival Probability

CONSULTA_CRT_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 85.0 mo    
% Excluding Normal Battery Depletion 100.0 99.7 99.3 98.9 98.8 98.7 98.7 98.7    
% Including Normal Battery Depletion 99.5 98.1 92.6 79.8 56.6 28.1 15.7 11.4    
#   58004 52868 45932 35409 20008 6804 531 116    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type CRT-Defib
NBD Code DDE-DDDR
Max. Delivered Energy 35 J
More
Connector Style DF4 CRT-D Style
Serial Number Prefix PUC
Mass 73 g
Volume 40 cc
Waveform Biphasic
X-Ray ID PUG
Hide this content
DF4 CRT-D Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.2 3.8 4.4 4.7
Quarterly 4.4 5.5 6.8 7.5
Semi-annual 4.8 6.2 7.9 9.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Integrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2012-01-09
CE Approval Date
Registered USA Implants2098
Estimated Active USA Implants1187
Normal Battery Depletions326
US Market Release2012-01-09
CE Approval Date
Estimated WW Distribution2190
Normal Battery Depletions326
Information

Malfunctions (USA)

Therapy Function Not Compromised 3
Battery 1
Electrical Component 1
Possible Early Battery Depletion 1
Therapy Function Compromised 0
Information

Malfunctions (WW)

Therapy Function Not Compromised 3
Battery 1
Electrical Component 1
Possible Early Battery Depletion 1
Therapy Function Compromised 0
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurments.

Information

Charge Time

D204TRM_D214TRM_D224TRK_D234TRK_CHRG
Loading...

Data as of July 31, 2017

www.medtronic.com/productperformance 501050