D154VWC Virtuoso VR
Implantable Cardioverter Defibrillator (ICD)
Non-Advisory Population
D154VWC Virtuoso VR
Information

Device Survival Probability


D154VWC,D164VWC_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 74 mo      
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.8 99.4 99.4 99.4      
% Including Normal Battery Depletion 99.9 99.8 99.6 99.2 98.1 94.2 93.6      
#   30854 28278 25609 20715 10977 1687 489      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVEV
Max. Delivered Energy 35J
More
Connector Style Cx Style
Serial Number Prefix PUN
Mass 68g
Volume 37cc
Waveform Biphasic
X-Ray ID PVR
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.8 5.1 5.3 5.4
Quarterly 8.1 9.0 9.6 10.0
Semi-annual 10.0 11.2 12.3 12.9
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

EOS device status is defined as 3 months following an RRT indication assuming the following PSP conditions: 100% VVI pacing at 60 bpm, 2.5 V, 0.4 ms; 500 Ohm pacing load; and 6 full-energy charges. EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Distribution Data

US Market Release2006-05-12
CE Approval Date
Registered USA Implants33111
Estimated Active USA Implants21610
Normal Battery Depletions233
US Market Release2006-05-12
CE Approval Date
Estimated WW Distribution33612
Normal Battery Depletions239
Information

Malfunctions (USA)

Therapy Function Not Compromised 121
Electrical Component 94
Electrical Interconnect 1
Possible Early Battery Depletion 12
Other 14
Therapy Function Compromised 12
Electrical Component 12
Information

Malfunctions (WW)

Therapy Function Not Compromised 125
Electrical Component 98
Electrical Interconnect 1
Possible Early Battery Depletion 12
Other 14
Therapy Function Compromised 12
Electrical Component 12
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.62v.

Information

Charge Time


D154VWC_D164VWC_CHRG

Data as of March 12, 2013