D154VWC Virtuoso VR
Implantable Cardioverter Defibrillator (ICD)
D154VWC Virtuoso VR
Information

Device Survival Probability

D154VWC,D164VWC_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 133.0 mo                  
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.7 98.1 96.8 96.8 96.7 96.6 96.6 96.6 96.6                  
% Including Normal Battery Depletion 99.5 99.3 99.1 98.5 95.6 86.5 75.0 56.1 39.1 26.1 11.9 9.7                  
#   28416 25901 23591 21569 19165 16021 12928 8902 5552 3197 402 208                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35 J
More
Connector Style Cx Style
Serial Number Prefix PUN
Mass 68g
Volume 37cc
Waveform Biphasic
X-Ray ID PVR
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.8 5.1 5.3 5.4
Quarterly 8.1 9.0 9.6 10.0
Semi-annual 10.0 11.2 12.3 12.9
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2006-05-12
CE Approval Date
Registered USA Implants33072
Estimated Active USA Implants6049
Normal Battery Depletions7497
US Market Release2006-05-12
CE Approval Date
Estimated WW Distribution33588
Normal Battery Depletions7519
Information

Malfunctions (USA)

Therapy Function Not Compromised 668
Battery 12
Electrical Component 636
Electrical Interconnect 1
Possible Early Battery Depletion 15
Other 4
Therapy Function Compromised 22
Battery 5
Electrical Component 17
Information

Malfunctions (WW)

Therapy Function Not Compromised 672
Battery 12
Electrical Component 640
Electrical Interconnect 1
Possible Early Battery Depletion 15
Other 4
Therapy Function Compromised 22
Battery 5
Electrical Component 17
Information

EOS Indication

EOS device status is defined as 3 months following an RRT indication assuming the following PSP conditions: 100% VVI pacing at 60 bpm, 2.5 V, 0.4 ms; 500 Ohm pacing load; and 6 full-energy charges. EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.62v.

Information

Charge Time

D154VWC_D164VWC_CHRG
Loading...

Data as of November 8, 2019

www.medtronic.com/productperformance 726050