D154VRC Entrust VR
Implantable Cardioverter Defibrillator (ICD)
D154VRC Entrust VR
Information

Device Survival Probability

D144VRC_D154VRC_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 137.0 mo                  
% Excluding Normal Battery Depletion 99.9 99.9 99.8 99.7 99.4 99.1 98.8 98.7 98.6 98.3 97.4 97.1                  
% Including Normal Battery Depletion 99.4 99.0 98.7 98.2 97.2 91.5 84.1 75.3 59.9 35.9 13.3 5.2                  
#   12605 11408 10198 8981 7899 6866 5846 4894 3535 1866 521 149                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35 J
More
Connector Style Cx Style
Serial Number Prefix PNT
Mass 68g
Volume 35cc
Waveform Biphasic
X-Ray ID PNT
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.8 5.0 5.2 5.3
Quarterly 7.5 8.3 8.8 9.0
Semi-annual 9.0 10.0 10.7 11.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2005-06-30
CE Approval Date2005-02-04
Registered USA Implants14465
Estimated Active USA Implants976
Normal Battery Depletions3165
US Market Release2005-06-30
CE Approval Date2005-02-04
Estimated WW Distribution23649
Normal Battery Depletions3429
Information

Malfunctions (USA)

Therapy Function Not Compromised 102
Battery 19
Electrical Component 47
Possible Early Battery Depletion 24
Other 12
Therapy Function Compromised 50
Battery 19
Electrical Component 27
Other 4
Information

Malfunctions (WW)

Therapy Function Not Compromised 144
Battery 21
Electrical Component 77
Electrical Interconnect 1
Possible Early Battery Depletion 30
Software/Firmware 1
Other 14
Therapy Function Compromised 62
Battery 23
Electrical Component 35
Other 4
Information

EOS Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 min-1, 2.5 V, 0.4 ms; 500 Ohm pacing load; and six full-energy charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.61v.

Information

Charge Time

D144VRC_D154VRC_CHRG
Loading...

Data as of April 11, 2019

www.medtronic.com/productperformance 651050