D154VRC
Entrust VR

Device Survival Probability
D144VRC_D154VRC_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | at 138.0 mo | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.9 | 99.9 | 99.8 | 99.7 | 99.4 | 99.1 | 98.8 | 98.7 | 98.6 | 98.3 | 97.4 | 97.1 | |||||||||
% | ![]() |
99.4 | 99.0 | 98.7 | 98.2 | 97.2 | 91.5 | 84.0 | 75.2 | 59.8 | 36.5 | 14.4 | 5.6 | |||||||||
# | 12605 | 11408 | 10197 | 8980 | 7898 | 6867 | 5847 | 4902 | 3583 | 1966 | 565 | 143 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Performance Notes For This Model
- There are no performance notes for this model.

Product Characteristics
Generator Type | ICD |
---|---|
NBD Code | VVE-VVIR |
Max. Delivered Energy | 35 J |
Connector Style | Cx Style |
---|---|
Serial Number Prefix | PNT |
Mass | 68g |
Volume | 35cc |
Waveform | Biphasic |
X-Ray ID | PNT |

Cx Style

Estimated Longevity
Charging Frequency | 100% Pacing | 50% Pacing | 15% Pacing | 100% Sensing |
---|---|---|---|---|
Monthly | 4.8 | 5.0 | 5.2 | 5.3 |
Quarterly | 7.5 | 8.3 | 8.8 | 9.0 |
Semi-annual | 9.0 | 10.0 | 10.7 | 11.0 |
Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Distribution Data
US Market Release | 2005-06-30 |
---|---|
CE Approval Date | 2005-02-04 |
Registered USA Implants | 14466 |
Estimated Active USA Implants | 995 |
Normal Battery Depletions | 3345 |
US Market Release | 2005-06-30 |
---|---|
CE Approval Date | 2005-02-04 |
Estimated WW Distribution | 23650 |
Normal Battery Depletions | 3653 |

EOS Indication
EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 min-1, 2.5 V, 0.4 ms; 500 Ohm pacing load; and six full-energy charges. EOL may be indicated before the end of three months if the device exceeds these conditions.
Data as of October 12, 2018