D154VRC Entrust VR
Implantable Cardioverter Defibrillator (ICD)
D154VRC Entrust VR
Information

Device Survival Probability


D144VRC_D154VRC_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 88 mo    
% Excluding Normal Battery Depletion 100.0 99.9 99.8 99.7 99.4 99.2 98.9 98.8    
% Including Normal Battery Depletion 99.9 99.6 99.4 98.9 98.0 92.5 85.4 82.1    
#   13657 12420 11154 9889 8434 6406 1457 176    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVEV
Max. Delivered Energy 35J
More
Connector Style Cx Style
Serial Number Prefix PNT
Mass 68g
Volume 35cc
Waveform Biphasic
X-Ray ID PNT
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.8 5.0 5.2 5.3
Quarterly 7.5 8.3 8.8 9.0
Semi-annual 9.0 10.0 10.7 11.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 min-1, 2.5 V, 0.4 ms; 500 Ohm pacing load; and six full-energy charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Distribution Data

US Market Release2005-06-30
CE Approval Date2005-02-04
Registered USA Implants14460
Estimated Active USA Implants7248
Normal Battery Depletions509
US Market Release2005-06-30
CE Approval Date2005-02-04
Estimated WW Distribution23681
Normal Battery Depletions662
Information

Malfunctions (USA)

Therapy Function Not Compromised 78
Battery 2
Electrical Component 42
Possible Early Battery Depletion 24
Other 10
Therapy Function Compromised 16
Electrical Component 16
Information

Malfunctions (WW)

Therapy Function Not Compromised 127
Battery 3
Electrical Component 72
Electrical Interconnect 1
Possible Early Battery Depletion 30
Software/Firmware 1
Other 20
Therapy Function Compromised 25
Battery 1
Electrical Component 24
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.61v.

Information

Charge Time


D144VRC_D154VRC_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050