D154VRC Entrust VR
Implantable Cardioverter Defibrillator (ICD)
D154VRC Entrust VR
Information

Device Survival Probability

D144VRC_D154VRC_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 138.0 mo                  
% Excluding Normal Battery Depletion 99.9 99.9 99.8 99.7 99.4 99.1 98.8 98.7 98.6 98.3 97.4 97.1                  
% Including Normal Battery Depletion 99.4 99.0 98.7 98.2 97.2 91.5 84.0 75.2 59.8 36.5 14.4 5.6                  
#   12605 11408 10197 8980 7898 6867 5847 4902 3583 1966 565 143                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35 J
More
Connector Style Cx Style
Serial Number Prefix PNT
Mass 68g
Volume 35cc
Waveform Biphasic
X-Ray ID PNT
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.8 5.0 5.2 5.3
Quarterly 7.5 8.3 8.8 9.0
Semi-annual 9.0 10.0 10.7 11.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2005-06-30
CE Approval Date2005-02-04
Registered USA Implants14466
Estimated Active USA Implants995
Normal Battery Depletions3345
US Market Release2005-06-30
CE Approval Date2005-02-04
Estimated WW Distribution23650
Normal Battery Depletions3653
Information

Malfunctions (USA)

Therapy Function Not Compromised 101
Battery 19
Electrical Component 47
Possible Early Battery Depletion 24
Other 11
Therapy Function Compromised 48
Battery 17
Electrical Component 27
Other 4
Information

Malfunctions (WW)

Therapy Function Not Compromised 142
Battery 20
Electrical Component 77
Electrical Interconnect 1
Possible Early Battery Depletion 30
Software/Firmware 1
Other 13
Therapy Function Compromised 60
Battery 21
Electrical Component 35
Other 4
Information

EOS Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 min-1, 2.5 V, 0.4 ms; 500 Ohm pacing load; and six full-energy charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.61v.

Information

Charge Time

D144VRC_D154VRC_CHRG
Loading...

Data as of October 12, 2018

www.medtronic.com/productperformance 626050