D154DRG EnTrust
Implantable Cardioverter Defibrillator (ICD)
D154DRG EnTrust
Information

Device Survival Probability


D144DRG_D154ATG_D154DRG_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 83 mo      
% Excluding Normal Battery Depletion 100.0 99.9 99.8 99.7 99.4 99.4 99.3      
% Including Normal Battery Depletion 99.8 99.6 99.1 97.9 90.7 70.2 34.1      
#   26329 24093 21667 19101 14919 9111 787      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code DDED, VVED
Max. Delivered Energy 35 J
More
Connector Style DR Style
Serial Number Prefix PNR
Mass 68 g
Volume 35 cc
Waveform Biphasic
X-Ray ID PNR
Hide this content
DR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.8 4.2 4.4 4.6
Quarterly 5.5 6.1 6.8 7.0
Semi-annual 6.1 7.0 7.9 8.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

3 months after RRT or > 16-second charge time

Information

Distribution Data

US Market Release
CE Approval Date2005-02-04
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release
CE Approval Date2005-02-04
Estimated WW Distribution
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Recommended Replacement (RRT)

Battery Voltage ≤ 2.61 V

Information

Charge Time


D144DRG_D154ATG_D154DRG_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050