D154AWG Virtuoso DR
Implantable Cardioverter Defibrillator (ICD)
D154AWG Virtuoso DR
Information

Device Survival Probability

D154AWG_D16AWG_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 125.0 mo                    
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.4 97.1 96.9 96.9 96.8 96.8 96.8 96.7                    
% Including Normal Battery Depletion 99.4 99.1 98.7 97.2 88.9 69.7 43.0 23.1 11.2 6.5 5.4                    
#   63126 57878 52707 47865 40635 29481 16331 7506 2982 1216 248                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code DDE-DDDR
Max. Delivered Energy 35 J
More
Connector Style DR Style
Serial Number Prefix PUL
Mass 68g
Volume 37cc
Waveform Biphasic
X-Ray ID PVR
Hide this content
DR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.1 4.5 4.8 5.0
Quarterly 6.3 7.3 8.3 8.8
Semi-annual 7.3 8.7 10.1 11.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2006-05-12
CE Approval Date
Registered USA Implants72611
Estimated Active USA Implants9558
Normal Battery Depletions20868
US Market Release2006-05-12
CE Approval Date
Estimated WW Distribution73345
Normal Battery Depletions20900
Information

Malfunctions (USA)

Therapy Function Not Compromised 1436
Battery 9
Electrical Component 1287
Electrical Interconnect 2
Possible Early Battery Depletion 132
Software/Firmware 3
Other 3
Therapy Function Compromised 35
Battery 3
Electrical Component 29
Possible Early Battery Depletion 1
Other 2
Information

Malfunctions (WW)

Therapy Function Not Compromised 1443
Battery 9
Electrical Component 1293
Electrical Interconnect 2
Possible Early Battery Depletion 132
Software/Firmware 4
Other 3
Therapy Function Compromised 35
Battery 3
Electrical Component 29
Possible Early Battery Depletion 1
Other 2
Information

EOS Indication

EOS device status is defined as 3 months following an RRT indication assuming the following PSP conditions: 100% DDD pacing at 60 bpm, 2.5 V, 0.4 ms; 500 Ohm pacing load; and 6 full-energy charges. EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.62v.

Information

Charge Time

D154AWG_D164AWG_CHRG
Loading...

Data as of November 8, 2019