D154ATG Entrust AT
Implantable Cardioverter Defibrillator (ICD)
D154ATG Entrust AT
Information

Device Survival Probability


D144DRG_D154ATG_D154DRG_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 83 mo      
% Excluding Normal Battery Depletion 100.0 99.9 99.8 99.7 99.4 99.4 99.3      
% Including Normal Battery Depletion 99.8 99.6 99.1 97.9 90.7 70.2 34.1      
#   26329 24093 21667 19101 14919 9111 787      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code DDED, VVED
Max. Delivered Energy 35J
More
Connector Style DR Style
Serial Number Prefix PNR
Mass 68g
Volume 35cc
Waveform Biphasic
X-Ray ID PNR
Hide this content
DR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.8 4.2 4.4 4.6
Quarterly 5.5 6.1 6.8 7.0
Semi-annual 6.1 7.0 7.9 8.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% DDD pacing at 60 min-1, 2.5 V, 0.4 ms; 500 Ohm pacing load; and six full-energy charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Distribution Data

US Market Release2005-06-30
CE Approval Date2005-02-04
Registered USA Implants28166
Estimated Active USA Implants9044
Normal Battery Depletions4524
US Market Release2005-06-30
CE Approval Date2005-02-04
Estimated WW Distribution37797
Normal Battery Depletions5015
Information

Malfunctions (USA)

Therapy Function Not Compromised 111
Electrical Component 28
Electrical Interconnect 1
Possible Early Battery Depletion 75
Software/Firmware 3
Other 4
Therapy Function Compromised 14
Electrical Component 14
Information

Malfunctions (WW)

Therapy Function Not Compromised 184
Electrical Component 76
Electrical Interconnect 3
Possible Early Battery Depletion 90
Software/Firmware 4
Other 11
Therapy Function Compromised 21
Battery 2
Electrical Component 19
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.61V.

Information

Charge Time


D144DRG_D154ATG_D154DRG_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050