Device Survival Probability
D153ATG_D153DRG_SURV
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | at 65.0 mo | ||||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.8 | 99.8 | 99.2 | 98.3 | 97.6 | 97.6 | ||||
% | ![]() |
98.2 | 98.2 | 97.1 | 91.1 | 66.6 | 42.2 | ||||
# | 408 | 374 | 337 | 276 | 192 | 105 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | ICD |
---|---|
NBD Code | DDE-DDDR |
Max. Delivered Energy | 30J |
Connector Style | DR Style |
---|---|
Serial Number Prefix | PNR |
Mass | 63g |
Volume | 33cc |
Waveform | Biphasic |
X-Ray ID | PNR |


Estimated Longevity
Charging Frequency | 100% Pacing | 50% Pacing | 15% Pacing | 100% Sensing |
---|---|---|---|---|
Monthly | 3.5 | 3.8 | 4.1 | 4.2 |
Quarterly | 4.8 | 5.4 | 6.0 | 6.3 |
Semi-annual | 5.3 | 6.1 | 6.9 | 7.2 |
Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Distribution Data
US Market Release | 2005-06-30 |
---|---|
CE Approval Date | |
Registered USA Implants | 459 |
Estimated Active USA Implants | 12 |
Normal Battery Depletions | 179 |
US Market Release | 2005-06-30 |
---|---|
CE Approval Date | |
Estimated WW Distribution | 598 |
Normal Battery Depletions | 184 |

EOS Indication
EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% DDD pacing at 60 bpm, 2.5 V, 0.4 ms; 500 Ohm pacing load; and six full-energy charges. EOL may be indicated before the end of three months if the device exceeds these conditions.
Data as of November 8, 2019