C3TR01 Consulta CRT-P
Cardiac Resynchronization Therapy - Pacing (CRT-Pacing)
C3TR01 Consulta CRT-P
Information

Device Survival Probability


C2TR01_C3TR01_C4TR01_SURV
    1 yr 2 yr                
% Excluding Normal Battery Depletion 100.0 99.9                
% Including Normal Battery Depletion 99.9 99.8                
#   6104 189                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type CRT-Pacing
NBD Code OAE-DDDR
Max. Delivered Energy N/A
More
Connector Style CRT-P Style
Serial Number Prefix PZI
Mass 26 g
Volume 15 cc
Waveform N/A
X-Ray ID PVX
Hide this content
CRT-P Style
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 8.7 10.7
Nominal 3.5 V (A, RV) 6.0 8.2
High 5.0 V (A, RV) 3.3 5.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Left Ventricle Pulse Width 0.4 ms; Left Ventricle Pulse Amplitude same as A, RV; Left Ventricle Percent Paced 100 %; Left Ventricle Lead Impedance same as A, RV; EGM Prestorage ON ; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content

Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release
CE Approval Date
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release
CE Approval Date
Estimated WW Distribution10335
Normal Battery Depletions6
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 1
Other 1
Therapy Function Compromised 0
Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.77V on 3 consecutive daily automatic measurments.

Data as of March 12, 2013

www.medtronic.com/productperformance 251050