ATDR01 Attesta DR MRI
Implantable Pulse Generator (IPG)
ATDR01 Attesta DR MRI
Information

Device Survival Probability

ATTESTA_DR_SURV
Loading...










    at 8.0 mo                  
% Excluding Normal Battery Depletion 100.0                  
% Including Normal Battery Depletion 100.0                  
#   108                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix FNB
Mass 27.1 g
Volume 12.1 cc
Waveform N/A
X-Ray ID PHX
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.5 10.8
Nominal 3.5 V (A, RV) 7.4 9.5
High 5.0 V (A, RV) 5.2 7.3
Footnote

Longevity estimates based on the following device usage.   Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 % Hide this content

Information

Distribution Data

US Market Release2017-08-03
CE Approval Date2017-06-16
Registered USA Implants448
Estimated Active USA Implants443
Normal Battery Depletions
US Market Release2017-08-03
CE Approval Date2017-06-16
Estimated WW Distribution67349
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 3
Electrical Component 3
Therapy Function Compromised 0
Information

EOL Indicator

180 days from RRT

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.5 V, or battery voltage less than or equal to 2.59 V and battery impedance at 3000 ohms

Data as of December 16, 2021

www.medtronic.com/productperformance 851050