ADSR06
Adapta SR
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | at 76 mo | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | ||||
| % | 100.0 | 99.9 | 99.8 | 99.4 | 98.7 | 97.4 | 97.3 | ||||
| # | 52770 | 38806 | 27047 | 16982 | 8557 | 1961 | 395 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
| Generator Type | IPG |
|---|---|
| NBD Code | SSIR |
| Max. Delivered Energy | N/A |
| Connector Style | 5/6mm UNI |
|---|---|
| Serial Number Prefix | NWN, PWN |
| Mass | 22.5 g |
| Volume | 11.0 cc |
| Waveform | N/A |
| X-Ray ID | PWB, NWN |
5/6mm UNI
Estimated Longevity
| Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
|---|---|---|
| Low 2.5 V (A, RV) | 7.4 | 8.9* | 7.9 | 9.7* |
| Nominal 3.5 V (A, RV) | 6.5 | 7.7* | 7.5 | 9.1* |
| High 5.0 V (A, RV) | 5.1 | 5.9* | 6.3 | 7.4* |
*The second longevity numbers provided apply to the N-series Adapta ADSR06 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 ; Hide this content
EOS Indication
From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.
Data as of March 12, 2013