ADDRS1 Adapta S DR
Implantable Pulse Generator (IPG)
ADDRS1 Adapta S DR
Information

Device Survival Probability

ADDRS1_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 125.0 mo                    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                    
% Including Normal Battery Depletion 99.6 99.5 99.3 98.7 97.3 91.7 78.0 61.6 48.9 40.6 35.5                    
#   41468 37863 34214 30541 26455 20609 13622 7414 3314 754 104                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PWA, NWA
Mass 23.6 g
Volume 11.1 cc
Waveform N/A
X-Ray ID PWB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 5.5 | 6.9* 6.1 | 7.8*
Nominal 3.5 V (A, RV) 4.3 | 5.3* 5.4 | 6.8*
High 5.0 V (A, RV) 3.2 | 3.7* 4.4 | 5.2*
Footnote

*The second longevity numbers provided apply to the N-series Adapta ADDRS1 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %; Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants48862
Estimated Active USA Implants25175
Normal Battery Depletions4459
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution83586
Normal Battery Depletions4490
Information

Malfunctions (USA)

Therapy Function Not Compromised 8
Electrical Component 5
Possible Early Battery Depletion 3
Therapy Function Compromised 6
Electrical Component 4
Other 2
Information

Malfunctions (WW)

Therapy Function Not Compromised 12
Electrical Component 9
Possible Early Battery Depletion 3
Therapy Function Compromised 10
Electrical Component 8
Other 2
Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of November 8, 2019

www.medtronic.com/productperformance 726050