ADDRL1 Adapta L DR
Implantable Pulse Generator (IPG)
ADDRL1 Adapta L DR
Information

Device Survival Probability

ADDRL1_SEDRL1_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 123.0 mo                    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                    
% Including Normal Battery Depletion 100.0 99.9 99.9 99.8 99.7 99.5 99.2 98.7 97.3 85.0 65.2                    
#   122549 111373 97653 82459 64737 47531 31928 19271 8980 1836 467                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

Performance Notes For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PWE, NWE
Mass 31.3 g
Volume 13.1 cc
Waveform N/A
X-Ray ID PWB, NWE
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.0 | 11.1* 10.0 | 12.6*
Nominal 3.5 V (A, RV) 7.3 | 8.7* 8.9 | 11.0*
High 5.0 V (A, RV) 5.4 | 6.1* 7.2 | 8.5*
Footnote

*The second longevity numbers provided apply to the N-series Adapta ADDRL1 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 % Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants137616
Estimated Active USA Implants110312
Normal Battery Depletions1163
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution291159
Normal Battery Depletions1691
Information

Malfunctions (USA)

Therapy Function Not Compromised 12
Electrical Component 10
Electrical Interconnect 1
Possible Early Battery Depletion 1
Therapy Function Compromised 4
Electrical Component 1
Electrical Interconnect 1
Other 2
Information

Malfunctions (WW)

Therapy Function Not Compromised 36
Electrical Component 33
Electrical Interconnect 2
Possible Early Battery Depletion 1
Therapy Function Compromised 15
Electrical Component 12
Electrical Interconnect 1
Other 2
Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of October 12, 2018