ADDRL1 Adapta DR
Implantable Pulse Generator (IPG)
ADDRL1 Adapta DR
Information

Device Survival Probability


ADDRL1_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 76 mo      
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0      
% Including Normal Battery Depletion 100.0 100.0 99.9 99.8 99.6 99.4 99.4      
#   56833 39040 23971 12383 4993 837 170      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PWE, NWE
Mass 31.3 g
Volume 13.1 cc
Waveform N/A
X-Ray ID PWB, NWE
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.0 | 11.1* 10.0 | 12.6*
Nominal 3.5 V (A, RV) 7.3 | 8.7* 8.9 | 11.0*
High 5.0 V (A, RV) 5.4 | 6.1* 7.2 | 8.5*
Footnote

*The second longevity numbers provided apply to the N-series Adapta ADDRL1 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 % Hide this content

Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants73888
Estimated Active USA Implants66688
Normal Battery Depletions23
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution177707
Normal Battery Depletions27
Information

Malfunctions (USA)

Therapy Function Not Compromised 9
Electrical Component 5
Electrical Interconnect 1
Other 3
Therapy Function Compromised 3
Electrical Interconnect 1
Other 2
Information

Malfunctions (WW)

Therapy Function Not Compromised 33
Electrical Component 27
Electrical Interconnect 2
Other 4
Therapy Function Compromised 10
Electrical Component 7
Electrical Interconnect 1
Other 2
Information

Recommended Replacement (RRT)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050