ADD01 Adapta D
Implantable Pulse Generator (IPG)
Advisories For This Model
- There are no advisories for this model.
Performance Notes For This Model
Product Characteristics
| Generator Type |
IPG |
| NBD Code |
DDD |
| Max. Delivered Energy |
N/A |
More
| Connector Style |
IS-1 UNI/BI |
| Serial Number Prefix |
PWF, NWF |
| Mass |
27.1 g |
| Volume |
12.1 cc |
| Waveform |
N/A |
| X-Ray ID |
PWB, NWF |
Hide this content
Estimated Longevity
| Amplitude Setting |
500 Lead Ω |
1000 Lead Ω |
| Low 2.5 V (A, RV) |
7.4 | 9.1* |
8.2 | 10.3* |
| Nominal 3.5 V (A, RV) |
6.0 | 7.2* |
7.4 | 9.0* |
| High 5.0 V (A, RV) |
4.5 | 5.1* |
6.0 | 7.0* |
EOS Indication
The 3 month prolonged service period between RRT/ERI and EOS
assumes 100% pacing in VVI mode, rate of 65 ppm, RV Amplitude
of 3.5 V, RV Pulse Width of 0.4 ms, and pacing load of 500 OHMS;.
Reprogramming the pacing parameters may reduce the duration of the
prolonged service period.
Recommended Replacement (RRT)
Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).
Data as of March 12, 2013