A3SR01
Advisa SR MRI
Device Survival Probability
A3SR01_ENSR01_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | at 90.0 mo | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | 100.0 | |||
% | 99.9 | 99.9 | 99.8 | 99.7 | 99.6 | 99.5 | 99.2 | 99.2 | |||
# | 22054 | 19406 | 17207 | 15022 | 12897 | 9606 | 3250 | 417 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | VVIR |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | PTR |
Mass | 21 g |
Volume | 11.9 cc |
Waveform | N/A |
X-Ray ID | PVX |
IS-1 UNI/BI
Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 11.9 | 12.9 |
Nominal 3.5 V (A, RV) | 10.0 | 11.6 |
High 5.0 V (A, RV) | 7.8 | 9.8 |
Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 ; Hide this content
Distribution Data
US Market Release | 2015-03-19 |
---|---|
CE Approval Date | 2014-04-24 |
Registered USA Implants | 28081 |
Estimated Active USA Implants | 16033 |
Normal Battery Depletions | 45 |
US Market Release | 2015-03-19 |
---|---|
CE Approval Date | 2014-04-24 |
Estimated WW Distribution | 85487 |
Normal Battery Depletions | 46 |
EOS Indication
The Prolonged Service Period (PSP) is the time between the RRT and EOS. The PSP is defined as 6 months assuming the following conditions: 100% VVI pacing at 60 min–1, 2.5 V RV pacing amplitude; 0.4 ms pulse width; and 600 Ω pacing load. The EOS may be indicated before the end of 6 months if the device exceeds these conditions.
Data as of December 1, 2023