A3DR01 Advisa DR MRI
Implantable Pulse Generator (IPG)
Advisories For This Model
- There are no advisories for this model.
Performance Notes For This Model
- There are no performance notes for this model.
Product Characteristics
Generator Type |
IPG |
NBD Code |
OAE-DDDR |
Max. Delivered Energy |
N/A |
More
Connector Style |
IS-1 UNI/BI |
Serial Number Prefix |
PZK |
Mass |
22g |
Volume |
12.7cc |
Waveform |
N/A |
X-Ray ID |
PVX |
Hide this content
Estimated Longevity
Amplitude Setting |
500 Lead Ω |
1000 Lead Ω |
Low 2.5 V (A, RV) |
8.9 |
10.2 |
Nominal 3.5 V (A, RV) |
6.9 |
8.6 |
High 5.0 V (A, RV) |
4.3 |
6.2 |
EOS Indication
From the point that the RRT is set, the pacemaker will operate for approximately six months for typical pacemaker configurations during the normal operating life.
Recommended Replacement (RRT)
Battery voltage less than or equal to 2.83V on 3 consecutive daily automatic measurments.
Data as of October 12, 2018