8158 Prodigy SR
Implantable Pulse Generator (IPG)
8158 Prodigy SR
Information

Device Survival Probability


8158_8160_8161_8162_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr at 173 mo            
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0            
% Including Normal Battery Depletion 99.8 99.6 99.2 98.6 97.8 96.7 95.0 91.9 87.1 77.1 66.1 55.1 46.1 42.0 38.7            
#   19684 17013 14714 12565 10706 9105 7661 6331 5010 3669 2526 1686 1136 433 110            
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PEM
Mass 22.5 g
Volume 11.2 cc
Waveform N/A
X-Ray ID PEM
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.8 10.7
Nominal 3.5 V (A, RV) 8.0 9.5
High 5.0 V (A, RV) 6.4 8.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Activity ON ; Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately 2.5 months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1996-07-01
CE Approval Date
Registered USA Implants4090
Estimated Active USA Implants640
Normal Battery Depletions317
US Market Release1996-07-01
CE Approval Date
Estimated WW Distribution4681
Normal Battery Depletions347
Information

Malfunctions (USA)

Therapy Function Not Compromised 2
Possible Early Battery Depletion 1
Other 1
Therapy Function Compromised 0
Information

Malfunctions (WW)

Therapy Function Not Compromised 5
Possible Early Battery Depletion 1
Other 4
Therapy Function Compromised 0
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050