8040 InSync
Cardiac Resynchronization Therapy - Pacing (CRT-Pacing)
8040 InSync
Information

Device Survival Probability


8040_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 121 mo                    
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.8 99.7 99.6 99.5 99.5 99.4 99.4 99.4                    
% Including Normal Battery Depletion 99.8 99.5 98.0 95.7 90.8 83.5 73.1 58.6 40.6 19.8 17.7                    
#   12258 10096 8105 6372 4972 3724 2618 1623 845 245 154                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type CRT-Pacing
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style CRT-P Style
Serial Number Prefix PIN
Mass 32 g
Volume 14.4 cc
Waveform N/A
X-Ray ID PIN
Hide this content
CRT-P Style
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 11.9 13.7
Nominal 3.5 V (A, RV) 8.9 11.4
High 5.0 V (A, RV) 6.6 9.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %; Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2001-08-28
CE Approval Date
Registered USA Implants15336
Estimated Active USA Implants1372
Normal Battery Depletions1346
US Market Release2001-08-28
CE Approval Date
Estimated WW Distribution20110
Normal Battery Depletions1382
Information

Malfunctions (USA)

Therapy Function Not Compromised 26
Electrical Component 4
Electrical Interconnect 16
Possible Early Battery Depletion 3
Other 3
Therapy Function Compromised 10
Electrical Interconnect 10
Information

Malfunctions (WW)

Therapy Function Not Compromised 27
Electrical Component 4
Electrical Interconnect 16
Possible Early Battery Depletion 3
Other 4
Therapy Function Compromised 11
Electrical Interconnect 11
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013