7304 InSync Maximo
Cardiac Resynchronization Therapy - Defibrillation (CRT-Defib)
7304 InSync Maximo
Information

Device Survival Probability


7304_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr at 62 mo        
% Excluding Normal Battery Depletion 100.0 99.9 99.6 99.2 99.1 99.1        
% Including Normal Battery Depletion 99.7 98.2 92.5 74.8 19.9 1.8        
#   16898 14754 11815 7487 1069 262        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type CRT-Defib
NBD Code VVED
Max. Delivered Energy 35 J
More
Connector Style CRT-D Style
Serial Number Prefix PRL
Mass 78 g
Volume 40 cc
Waveform Biphasic
X-Ray ID PRL
Hide this content
CRT-D Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.3 3.8 4.1 4.3
Quarterly 4.5 5.3 6.2 6.6
Semi-annual 5.0 6.0 7.1 7.7
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Intergrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2005-04-08
CE Approval Date2005-01-14
Registered USA Implants18953
Estimated Active USA Implants2000
Normal Battery Depletions5150
US Market Release2005-04-08
CE Approval Date2005-01-14
Estimated WW Distribution30906
Normal Battery Depletions5239
Information

Malfunctions (USA)

Therapy Function Not Compromised 106
Battery 1
Electrical Component 13
Possible Early Battery Depletion 89
Other 3
Therapy Function Compromised 5
Electrical Component 4
Other 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 143
Battery 1
Electrical Component 34
Possible Early Battery Depletion 100
Other 8
Therapy Function Compromised 11
Electrical Component 10
Other 1
Information

Elective Replacement (ERI)

Battery voltage less than 2.62 V or Charge Time greater than 16 seconds.

Information

Charge Time


7304_CHRG

Data as of March 12, 2013