7290Cx Onyx
Implantable Cardioverter Defibrillator (ICD)
7290Cx Onyx
Information

Device Survival Probability


7290_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 88 mo    
% Excluding Normal Battery Depletion 99.9 99.5 99.5 99.4 99.4 99.4 99.4 99.4    
% Including Normal Battery Depletion 99.8 99.0 98.4 97.2 93.4 82.9 60.5 48.2    
#   866 789 709 626 541 409 184 103    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVEV
Max. Delivered Energy 30J
More
Connector Style Cx Style
Serial Number Prefix PRP
Mass 77g
Volume 39cc
Waveform Biphasic
X-Ray ID PRP
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.8 4.1 4.3 4.5
Quarterly 5.0 5.6 6.2 6.4
Semi-annual 5.4 6.1 6.7 7.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOL Indication

Battery voltage less than or equal to 2.40V. The minimum time between ERI and End of Life is three months, assuming 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 ohm pacing load; and six 30 J charges.

Information

Distribution Data

US Market Release2004-03-08
CE Approval Date2004-05-04
Registered USA Implants954
Estimated Active USA Implants290
Normal Battery Depletions145
US Market Release2004-03-08
CE Approval Date2004-05-04
Estimated WW Distribution2024
Normal Battery Depletions164
Information

Malfunctions (USA)

Therapy Function Not Compromised 4
Electrical Component 3
Possible Early Battery Depletion 1
Therapy Function Compromised 1
Electrical Component 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 4
Electrical Component 3
Possible Early Battery Depletion 1
Therapy Function Compromised 1
Electrical Component 1
Information

Elective Replacement (ERI)

Battery voltage less than or equal to 2.55V.

Information

Charge Time


7290_CHRG

Data as of March 12, 2013