7232E Maximo VR
Implantable Cardioverter Defibrillator (ICD)
7232E Maximo VR
Information

Device Survival Probability

7232_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 153.0 mo                
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.8 99.8 99.8 99.8 99.8 99.8 99.8 99.8 99.8 99.8                
% Including Normal Battery Depletion 99.4 99.1 98.8 98.3 96.7 90.8 82.3 70.1 45.0 19.6 15.1 14.1 13.3                
#   38269 34244 30526 26920 23716 20620 17420 13957 8444 3139 1841 1070 115                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35J
More
Connector Style E Style
Serial Number Prefix PVG
Mass 76g
Volume 39cc
Waveform Biphasic
X-Ray ID PVG
Hide this content
E Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.4 4.7 4.8 4.9
Quarterly 7.0 7.5 8.0 8.3
Semi-annual 8.2 9.0 9.7 10.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2003-10-06
CE Approval Date2004-10-22
Registered USA Implants490
Estimated Active USA Implants68
Normal Battery Depletions88
US Market Release2003-10-06
CE Approval Date2004-10-22
Estimated WW Distribution580
Normal Battery Depletions90
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Component 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Component 1
Information

EOL Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 Ohm pacing load; and six 30 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Elective Replacement (ERI)

Battery voltage less than or equal to 2.62v or charge time greater than 16 seconds.

Information

Charge Time

7232_CHRG
Loading...

Data as of October 12, 2018

www.medtronic.com/productperformance 626050