7232Cx Maximo VR
Implantable Cardioverter Defibrillator (ICD)
7232Cx Maximo VR
Information

Device Survival Probability

7232_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr at 208.0 mo      
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.9 99.9 99.9 99.8 99.8 99.7 99.7 99.7 99.7 99.7 99.6 99.6 99.6 99.6 99.6      
% Including Normal Battery Depletion 99.8 99.7 99.3 98.9 97.5 92.7 85.8 75.4 52.4 27.3 22.4 21.2 20.1 19.4 18.4 17.6 16.6 16.0      
#   38514 35189 31911 28429 25002 21613 18239 14653 9037 3706 2514 2013 1574 1226 932 645 308 113      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35J
More
Connector Style Cx Style
Serial Number Prefix PRN
Mass 76g
Volume 39cc
Waveform Biphasic
X-Ray ID PRN
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.4 4.7 4.8 4.9
Quarterly 7.0 7.5 8.0 8.3
Semi-annual 8.2 9.0 9.7 10.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2003-10-06
CE Approval Date2003-10-28
Registered USA Implants43623
Estimated Active USA Implants2750
Normal Battery Depletions10395
US Market Release2003-10-06
CE Approval Date2003-10-28
Estimated WW Distribution75208
Normal Battery Depletions10702
Information

Malfunctions (USA)

Therapy Function Not Compromised 58
Electrical Component 29
Possible Early Battery Depletion 25
Software/Firmware 2
Other 2
Therapy Function Compromised 15
Electrical Component 12
Electrical Interconnect 1
Possible Early Battery Depletion 1
Other 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 115
Battery 2
Electrical Component 70
Possible Early Battery Depletion 35
Software/Firmware 5
Other 3
Therapy Function Compromised 23
Electrical Component 20
Electrical Interconnect 1
Possible Early Battery Depletion 1
Other 1
Information

EOL Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 Ohm pacing load; and six 30 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Elective Replacement (ERI)

Battery voltage less than or equal to 2.62v or charge time greater than 16 seconds.

Information

Charge Time

7232_CHRG
Loading...

Data as of December 1, 2023

www.medtronic.com/productperformance 951051