7231Cx GEM III VR
Implantable Cardioverter Defibrillator (ICD)
7231Cx GEM III VR
Information

Device Survival Probability


7231_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 106 mo  
% Excluding Normal Battery Depletion 99.9 99.9 99.8 99.8 99.8 99.7 99.7 99.7 99.7  
% Including Normal Battery Depletion 99.8 99.5 99.1 98.7 97.3 89.7 78.5 53.3 10.2  
#   15789 14113 12514 11028 9630 7910 5898 3024 353  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVEV
Max. Delivered Energy 30J
More
Connector Style Cx Style
Serial Number Prefix PJL
Mass 77g
Volume 39cc
Waveform Biphasic
X-Ray ID PJL
Hide this content
Cx Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.3 4.7 5.0 5.2
Quarterly 6.0 6.8 7.4 7.8
Semi-annual 6.6 7.5 8.5 8.9
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOL Indication

Battery voltage less than or equal to 2.40V. The minimum time between ERI and End of Life is three months, assuming 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 ohm pacing load; and six 30 J charges.

Information

Distribution Data

US Market Release2000-12-12
CE Approval Date2000-12-08
Registered USA Implants17448
Estimated Active USA Implants1884
Normal Battery Depletions3612
US Market Release2000-12-12
CE Approval Date2000-12-08
Estimated WW Distribution33330
Normal Battery Depletions3800
Information

Malfunctions (USA)

Therapy Function Not Compromised 30
Battery 1
Electrical Component 22
Possible Early Battery Depletion 4
Other 3
Therapy Function Compromised 10
Battery 1
Electrical Component 8
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 78
Battery 1
Electrical Component 57
Electrical Interconnect 1
Possible Early Battery Depletion 7
Software/Firmware 1
Other 11
Therapy Function Compromised 19
Battery 2
Electrical Component 16
Electrical Interconnect 1
Information

Elective Replacement (ERI)

Battery voltage less than or equal to 2.55V.

Information

Charge Time


7231_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050