7230E
Marquis VR
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | at 111 mo | ||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.9 | 99.9 | 99.9 | 99.8 | 99.7 | 99.6 | 99.5 | 99.4 | 99.3 | 99.3 | |
| % | 99.7 | 99.5 | 99.1 | 98.8 | 98.2 | 94.7 | 86.9 | 75.1 | 39.9 | 14.3 | |
| # | 17348 | 13575 | 11346 | 10171 | 9087 | 7917 | 6534 | 4708 | 950 | 179 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Product Characteristics
| Generator Type | ICD |
|---|---|
| NBD Code | VVEV |
| Max. Delivered Energy | 30J |
| Connector Style | E Style |
|---|---|
| Serial Number Prefix | PLY |
| Mass | 75g |
| Volume | 36cc |
| Waveform | Biphasic |
| X-Ray ID | PLY |
E Style
Estimated Longevity
| Charging Frequency | 100% Pacing | 50% Pacing | 15% Pacing | 100% Sensing |
|---|---|---|---|---|
| Monthly | 4.9 | 5.2 | 5.4 | 5.5 |
| Quarterly | 7.3 | 8.0 | 8.5 | 8.7 |
| Semi-annual | 8.5 | 9.3 | 10.0 | 10.4 |
Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content
EOL Indication
EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 Ohm pacing load; and six 30 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.
Data as of March 12, 2013