7230B Marquis VR
Implantable Cardioverter Defibrillator (ICD)
7230B Marquis VR
Information

Device Survival Probability


7230_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 111 mo
% Excluding Normal Battery Depletion 99.9 99.9 99.9 99.8 99.7 99.6 99.5 99.4 99.3 99.3
% Including Normal Battery Depletion 99.7 99.5 99.1 98.8 98.2 94.7 86.9 75.1 39.9 14.3
#   17348 13575 11346 10171 9087 7917 6534 4708 950 179
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVEV
Max. Delivered Energy 30J
More
Connector Style B Style
Serial Number Prefix PLW
Mass 75g
Volume 36cc
Waveform Biphasic
X-Ray ID PKD
Hide this content
B Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.9 5.2 5.4 5.5
Quarterly 7.3 8.0 8.5 8.7
Semi-annual 8.5 9.3 10.0 10.4
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOL Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 Ohm pacing load; and six 30 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Distribution Data

US Market Release2002-12-17
CE Approval Date2002-08-21
Registered USA Implants237
Estimated Active USA Implants21
Normal Battery Depletions17
US Market Release2002-12-17
CE Approval Date2002-08-21
Estimated WW Distribution285
Normal Battery Depletions18
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Battery 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 1
Other 1
Therapy Function Compromised 1
Battery 1
Information

Elective Replacement (ERI)

Battery voltage less than or equal to 2.62v or charge time greater than 16 seconds.

Information

Charge Time


7230_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050