6966 Transvene
Defibrillation
6966 Transvene
ALL
Information

Device Survival Probability


6936_6966_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 150 mo                
% Survival 98.2 97.0 95.2 94.3 91.1 86.8 79.8 75.9 70.2 65.8 63.6 61.8 59.6                
#   1140 956 807 659 530 414 284 210 142 104 75 56 50                

System Longevity Study Results

Number of Leads Enrolled in Study 1349
Cumulative Months of Follow-Up 75241
Number of Leads Active in Study 13
Information 188

Qualifying Complications

Conductor Fracture 22
Extra Cardiac Stimulation 6
Failure to Capture 15
Failure to Sense 7
Impedance Out of Range 7
Insulation (not further defined) 14
Oversensing 112
Unspecified Clinical Failure 5
Information

US Acute Lead Observations

Cardiac Perforation 2
Conductor Fracture 2
Failure to Capture 2
Failure to Sense 1
Insulation Breach 1
Lead Dislodgement 1
Unspecified Clinical Failure 3
Information

Distribution Data

US Market Release1993-12-09
CE Approval Date
Registered USA Implants4687
Estimated Active USA Implants310
US Market Release1993-12-09
CE Approval Date
Estimated WW Distribution5098
Information

USA Returned Product Analysis

Conductor Fracture 46
Insulation Breach 41
Crimp/Weld/Bond 0
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 54
Insulation Breach 44
Crimp/Weld/Bond 1
Other 15
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix TAL
Lead Body Diameter 10 FR
Insulation Polyurethane
Pace/Sense Connector Style 3.2 mm
High Voltage Connector 6.5 mm
Steroid None
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050