6947M Sprint Quattro Secure
Defibrillation
6947M Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947M_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 78.0 mo      
% Survival 99.7 99.5 99.3 99.3 99.0 99.0 97.7      
#   1715 1434 1229 980 782 416 182      

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2113
Cumulative Months of Follow-Up 91022
Number of Leads Active in Study 954
Information 17

Qualifying Complications

Conductor Fracture 9
Failure to Capture 4
Failure to Sense 2
Impedance Out of Range 0
Lead Dislodgement 1
Other 1
Information

US Acute Lead Observations

Cardiac Perforation 30
Conductor Fracture 9
Extra Cardiac Stimulation 10
Failure to Capture 94
Failure to Sense 34
Impedance Out of Range 26
Lead Dislodgement 193
Oversensing 69
Information

Distribution Data

US Market Release2012-02-13
CE Approval Date2010-03-12
Registered USA Implants115553
Estimated Active USA Implants103417
US Market Release2012-02-13
CE Approval Date2010-03-12
Estimated WW Distribution228460
Information

USA Returned Product Analysis

Conductor Fracture 145
Insulation Breach 13
Crimp/Weld/Bond 0
Other 23
Information

WW Returned Product Analysis

Conductor Fracture 254
Insulation Breach 19
Crimp/Weld/Bond 0
Other 27
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDK
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style DF4-LLHH
High Voltage Connector DF4-LLHH
Steroid Yes
Hide this content

Data as of November 8, 2019

www.medtronic.com/productperformance 726050