6947 Sprint Quattro Secure
Defibrillation
6947 Sprint Quattro Secure
ALL
Information

Device Survival Probability

6947_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr 18 yr 19 yr    
% Survival 99.5 99.3 99.0 98.7 98.2 97.9 97.5 97.1 96.7 96.1 95.8 95.4 94.9 94.3 93.9 91.5 91.5 90.7 88.2    
#   3297 2899 2546 2255 2022 1782 1533 1372 1223 1049 861 638 456 280 177 139 117 98 59    

Product Surveillance Registry Results

Number of Leads Enrolled in Study 4568
Cumulative Months of Follow-Up 294892
Number of Leads Active in Study 618
Information 102

Qualifying Complications

Cardiac Perforation 1
Conductor Fracture 39
Failure to Capture 8
Failure to Sense 2
Impedance Out of Range 13
Insulation (not further defined) 6
Lead Dislodgement 5
Oversensing 21
Other 4
Unspecified Clinical Failure 3
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 29
Conductor Fracture 26
Extra Cardiac Stimulation 2
Failure to Capture 83
Failure to Sense 36
Impedance Out of Range 61
Insulation Breach 4
Lead Dislodgement 124
Oversensing 141
Unspecified Clinical Failure 20
Information

Distribution Data

US Market Release2001-11-12
CE Approval Date2001-10-04
Registered USA Implants375502
Estimated Active USA Implants126103
US Market Release2001-11-12
CE Approval Date2001-10-04
Estimated WW Distribution616493
Information

USA Returned Product Analysis

Conductor Fracture 1398
Insulation Breach 103
Crimp/Weld/Bond 4
Other 197
Information

WW Returned Product Analysis

Conductor Fracture 1732
Insulation Breach 134
Crimp/Weld/Bond 4
Other 239
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/Two Coils
More
Serial Number Prefix TDG
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes