6947 Sprint Quattro Secure
Defibrillation
6947 Sprint Quattro Secure
ALL
Information

Device Survival Probability


6947_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117 mo
% Survival 99.5 99.2 98.9 98.6 98.2 97.8 97.3 96.9 95.8 95.8
#   2395 1900 1176 745 582 437 327 223 105 48

System Longevity Study Results

Number of Leads Enrolled in Study 2715
Cumulative Months of Follow-Up 115003
Number of Leads Active in Study 1151
Information 37

Qualifying Complications

Conductor Fracture 7
Failure to Capture 1
Failure to Sense 2
Impedance Out of Range 7
Insulation (not further defined) 3
Lead Dislodgement 4
Oversensing 11
Unspecified Clinical Failure 2
Information

US Acute Lead Observations

Cardiac Perforation 22
Conductor Fracture 18
Extra Cardiac Stimulation 2
Failure to Capture 66
Failure to Sense 28
Impedance Out of Range 45
Insulation Breach 4
Lead Dislodgement 99
Oversensing 109
Unspecified Clinical Failure 22
Information

Distribution Data

US Market Release2001-11-12
CE Approval Date2001-10-04
Registered USA Implants358933
Estimated Active USA Implants243668
US Market Release2001-11-12
CE Approval Date2001-10-04
Estimated WW Distribution469857
Information

USA Returned Product Analysis

Conductor Fracture 436
Insulation Breach 32
Crimp/Weld/Bond 4
Other 272
Information

WW Returned Product Analysis

Conductor Fracture 599
Insulation Breach 46
Crimp/Weld/Bond 4
Other 357
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix TDG
Lead Body Diameter 8.6 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013