6945 Sprint
Defibrillation
6945 Sprint
ALL
Information

Device Survival Probability


6945_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 141 mo                  
% Survival 99.4 98.7 98.2 97.6 96.8 96.1 95.6 94.5 93.2 92.7 91.7 91.7                  
#   997 822 684 566 466 382 315 239 183 137 86 49                  

System Longevity Study Results

Number of Leads Enrolled in Study 1155
Cumulative Months of Follow-Up 67487
Number of Leads Active in Study 125
Information 38

Qualifying Complications

Conductor Fracture 8
Extra Cardiac Stimulation 1
Failure to Capture 2
Failure to Sense 4
Impedance Out of Range 5
Oversensing 17
Unspecified Clinical Failure 1
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 1
Extra Cardiac Stimulation 1
Failure to Capture 6
Failure to Sense 2
Insulation Breach 2
Lead Dislodgement 4
Oversensing 7
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release1997-09-26
CE Approval Date
Registered USA Implants42716
Estimated Active USA Implants12821
US Market Release1997-09-26
CE Approval Date
Estimated WW Distribution51221
Information

USA Returned Product Analysis

Conductor Fracture 121
Insulation Breach 35
Crimp/Weld/Bond 1
Other 13
Information

WW Returned Product Analysis

Conductor Fracture 150
Insulation Breach 43
Crimp/Weld/Bond 1
Other 22
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Integrated Bipolar
More
Serial Number Prefix TDA
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013