6945 Sprint
Defibrillation
6945 Sprint
ALL
Information

Device Survival Probability

6945_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr              
% Survival 99.4 98.6 98.1 97.5 96.5 95.7 95.0 93.9 92.5 91.5 90.3 89.6 88.9 86.8              
#   1020 824 652 520 403 310 273 229 186 155 133 118 90 53              

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1192
Cumulative Months of Follow-Up 66751
Number of Leads Active in Study 87
Information 45

Qualifying Complications

Conductor Fracture 11
Extra Cardiac Stimulation 1
Failure to Capture 2
Failure to Sense 4
Impedance Out of Range 6
Oversensing 19
Other 1
Unspecified Clinical Failure 1
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 1
Extra Cardiac Stimulation 1
Failure to Capture 6
Failure to Sense 2
Impedance Out of Range 1
Insulation Breach 2
Lead Dislodgement 4
Oversensing 8
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release1997-09-26
CE Approval Date
Registered USA Implants42724
Estimated Active USA Implants9832
US Market Release1997-09-26
CE Approval Date
Estimated WW Distribution52533
Information

USA Returned Product Analysis

Conductor Fracture 146
Insulation Breach 45
Crimp/Weld/Bond 1
Other 6
Information

WW Returned Product Analysis

Conductor Fracture 173
Insulation Breach 54
Crimp/Weld/Bond 1
Other 14
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Integrated Bipolar/Two Coils
More
Serial Number Prefix TDA
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of April 14, 2016