6944 Sprint Quattro
Defibrillation
6944 Sprint Quattro
ALL
Information

Device Survival Probability


6944_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 81 mo      
% Survival 100.0 100.0 99.3 97.5 95.1 95.1 95.1      
#   356 217 128 85 66 58 50      

System Longevity Study Results

Number of Leads Enrolled in Study 509
Cumulative Months of Follow-Up 16123
Number of Leads Active in Study 283
Information 6

Qualifying Complications

Conductor Fracture 1
Failure to Sense 1
Impedance Out of Range 1
Oversensing 2
Unspecified Clinical Failure 1
Information

US Acute Lead Observations

Conductor Fracture 2
Failure to Capture 11
Failure to Sense 3
Impedance Out of Range 8
Lead Dislodgement 16
Oversensing 11
Unspecified Clinical Failure 6
Information

Distribution Data

US Market Release2000-12-13
CE Approval Date1999-11-05
Registered USA Implants41390
Estimated Active USA Implants21805
US Market Release2000-12-13
CE Approval Date1999-11-05
Estimated WW Distribution83674
Information

USA Returned Product Analysis

Conductor Fracture 93
Insulation Breach 2
Crimp/Weld/Bond 1
Other 31
Information

WW Returned Product Analysis

Conductor Fracture 219
Insulation Breach 8
Crimp/Weld/Bond 2
Other 62
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Tines
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix TDC
Lead Body Diameter 8.2 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013