6943 Sprint
Defibrillation
6943 Sprint
ALL
Information

Device Survival Probability


6943_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 135 mo                  
% Survival 98.6 97.7 96.5 95.6 93.7 92.1 90.9 89.9 87.8 86.6 83.1 83.1                  
#   1146 981 874 745 628 511 399 293 196 116 65 53                  

System Longevity Study Results

Number of Leads Enrolled in Study 1311
Cumulative Months of Follow-Up 81556
Number of Leads Active in Study 220
Information 94

Qualifying Complications

Conductor Fracture 24
Failure to Capture 8
Failure to Sense 7
Impedance Out of Range 8
Insulation (not further defined) 1
Lead Dislodgement 2
Oversensing 40
Other 1
Unspecified Clinical Failure 3
Information

US Acute Lead Observations

Cardiac Perforation 1
Failure to Capture 1
Failure to Sense 1
Impedance Out of Range 2
Insulation Breach 1
Oversensing 1
Information

Distribution Data

US Market Release1997-10-06
CE Approval Date
Registered USA Implants20481
Estimated Active USA Implants6308
US Market Release1997-10-06
CE Approval Date
Estimated WW Distribution31932
Information

USA Returned Product Analysis

Conductor Fracture 71
Insulation Breach 28
Crimp/Weld/Bond 1
Other 7
Information

WW Returned Product Analysis

Conductor Fracture 216
Insulation Breach 82
Crimp/Weld/Bond 1
Other 28
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix TCE
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013