6942 Sprint
Defibrillation
6942 Sprint
ALL
Information

Device Survival Probability


6942_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 111 mo
% Survival 99.1 99.1 98.1 97.5 96.8 96.8 96.8 96.8 96.8 96.8
#   294 231 179 141 118 100 78 67 53 49

System Longevity Study Results

Number of Leads Enrolled in Study 351
Cumulative Months of Follow-Up 19099
Number of Leads Active in Study 29
Information 7

Qualifying Complications

Conductor Fracture 1
Failure to Sense 1
Lead Dislodgement 1
Oversensing 3
Unspecified Clinical Failure 1
Information

US Acute Lead Observations

Conductor Fracture 1
Failure to Capture 4
Impedance Out of Range 2
Lead Dislodgement 1
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1997-07-18
CE Approval Date
Registered USA Implants17679
Estimated Active USA Implants5401
US Market Release1997-07-18
CE Approval Date
Estimated WW Distribution22980
Information

USA Returned Product Analysis

Conductor Fracture 14
Insulation Breach 22
Crimp/Weld/Bond 1
Other 7
Information

WW Returned Product Analysis

Conductor Fracture 21
Insulation Breach 25
Crimp/Weld/Bond 1
Other 12
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Tines
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Integrated Bipolar
More
Serial Number Prefix TCB
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013