6942
Sprint
ALL
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | at 111 mo | ||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.1 | 99.1 | 98.1 | 97.5 | 96.8 | 96.8 | 96.8 | 96.8 | 96.8 | 96.8 | |
| # | 294 | 231 | 179 | 141 | 118 | 100 | 78 | 67 | 53 | 49 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 351 |
|---|---|
| Cumulative Months of Follow-Up | 19099 |
| Number of Leads Active in Study | 29 |
US Acute Lead Observations
| Conductor Fracture | 1 |
|---|---|
| Failure to Capture | 4 |
| Impedance Out of Range | 2 |
| Lead Dislodgement | 1 |
| Unspecified Clinical Failure | 1 |
Product Characteristics
| Lead Type | Defibrillation |
|---|---|
| Type and/or Fixation | Tines |
| Lead Function | Pacing/Sensing and Defibrillation |
| Lead Placement | Transvenous |
| Lead Tip Location | Right Ventricle |
| Pace/Sense Polarity | Integrated Bipolar |
| Serial Number Prefix | TCB |
|---|---|
| Lead Body Diameter | 7.8 FR |
| Insulation | Silicone |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | DF-1 |
| Steroid | Yes |
Data as of March 12, 2013