6937A
SVC/CS
ALL
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | 10 yr | 11 yr | at 138 mo | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 98.4 | 97.5 | 97.2 | 96.7 | 95.4 | 94.9 | 93.9 | 93.4 | 92.2 | 91.2 | 91.2 | 91.2 | ||||||||||
| # | 809 | 677 | 566 | 470 | 381 | 298 | 213 | 163 | 108 | 77 | 57 | 50 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 966 |
|---|---|
| Cumulative Months of Follow-Up | 54382 |
| Number of Leads Active in Study | 13 |
Performance Notes For This Model
Product Characteristics
| Lead Type | Defibrillation |
|---|---|
| Type and/or Fixation | Passive |
| Lead Function | Defibrillation Electrode |
| Lead Placement | Transvenous |
| Lead Tip Location | SVC/CS |
| Pace/Sense Polarity | n/a |
| Serial Number Prefix | TDB |
|---|---|
| Lead Body Diameter | 7.5 FR |
| Insulation | Silicone with Polyurethane Overlay |
| Pace/Sense Connector Style | n/a |
| High Voltage Connector | DF-1 |
| Steroid | None |
Data as of March 12, 2013