6935M
Sprint Quattro Secure S
ALL
Device Survival Probability
6935M_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | 9 yr | at 114.0 mo | ||
---|---|---|---|---|---|---|---|---|---|---|---|
% | 99.6 | 99.5 | 99.2 | 98.9 | 98.6 | 98.2 | 97.7 | 97.4 | 97.0 | 97.0 | |
# | 6732 | 5312 | 4334 | 3508 | 2833 | 2131 | 1302 | 634 | 259 | 139 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 9312 |
---|---|
Cumulative Months of Follow-Up | 371545 |
Number of Leads Active in Study | 4439 |
US Acute Lead Observations
Cardiac Perforation | 183 |
---|---|
Conductor Fracture | 21 |
Extra Cardiac Stimulation | 31 |
Failure to Capture | 422 |
Failure to Sense | 146 |
Impedance Out of Range | 126 |
Insulation Breach | 3 |
Lead Dislodgement | 633 |
Oversensing | 327 |
Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/One Coil |
Serial Number Prefix | TDL |
---|---|
Lead Body Diameter | 2.8 mm (8.4 Fr) |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | DF4-LLHO |
High Voltage Connector | DF4-LLHO |
Steroid | Yes |
Data as of December 1, 2023