6935 Sprint Quattro Secure
Defibrillation
6935 Sprint Quattro Secure
ALL
Information

Device Survival Probability


6935_SURV
    1 yr 2 yr 3 yr at 39 mo            
% Survival 99.5 99.2 99.2 99.2            
#   846 353 81 44            

System Longevity Study Results

Number of Leads Enrolled in Study 1886
Cumulative Months of Follow-Up 25696
Number of Leads Active in Study 1610
Information 8

Qualifying Complications

Conductor Fracture 1
Failure to Sense 1
Lead Dislodgement 4
Oversensing 2
Information

US Acute Lead Observations

Cardiac Perforation 10
Failure to Capture 15
Failure to Sense 5
Impedance Out of Range 10
Insulation Breach 1
Lead Dislodgement 26
Oversensing 23
Unspecified Clinical Failure 5
Information

Distribution Data

US Market Release2008-11-01
CE Approval Date2008-03-31
Registered USA Implants40213
Estimated Active USA Implants36619
US Market Release2008-11-01
CE Approval Date2008-03-31
Estimated WW Distribution75996
Information

USA Returned Product Analysis

Conductor Fracture 46
Insulation Breach 2
Crimp/Weld/Bond 0
Other 42
Information

WW Returned Product Analysis

Conductor Fracture 119
Insulation Breach 4
Crimp/Weld/Bond 0
Other 74
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix TAU
Lead Body Diameter 8,2 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013