6935 Sprint Quattro Secure
Defibrillation
6935 Sprint Quattro Secure
ALL
Information

Device Survival Probability

6935_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr at 54.0 mo          
% Survival 99.4 99.2 98.9 98.2 98.2          
#   1872 1203 636 249 104          

Product Surveillance Registry Results

Number of Leads Enrolled in Study 2350
Cumulative Months of Follow-Up 61342
Number of Leads Active in Study 1438
Information 23

Qualifying Complications

Conductor Fracture 7
Failure to Capture 2
Failure to Sense 1
Impedance Out of Range 1
Lead Dislodgement 7
Oversensing 4
Other 1
Information

US Acute Lead Observations

Cardiac Perforation 14
Failure to Capture 19
Failure to Sense 5
Impedance Out of Range 14
Insulation Breach 1
Lead Dislodgement 33
Oversensing 37
Unspecified Clinical Failure 5
Information

Distribution Data

US Market Release2008-11-01
CE Approval Date2008-03-31
Registered USA Implants49968
Estimated Active USA Implants43965
US Market Release2008-11-01
CE Approval Date2008-03-31
Estimated WW Distribution97707
Information

USA Returned Product Analysis

Conductor Fracture 133
Insulation Breach 4
Crimp/Weld/Bond 0
Other 37
Information

WW Returned Product Analysis

Conductor Fracture 308
Insulation Breach 10
Crimp/Weld/Bond 0
Other 60
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar/One Coil
More
Serial Number Prefix TAU
Lead Body Diameter 8,2 FR
Insulation Silicone with Polyurethane Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes