6935
Sprint Quattro Secure S
ALL

Device Survival Probability
6935_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
99.4 | 99.2 | 98.9 | 98.6 | 98.4 | 98.0 | 96.9 | 96.1 | ||
# | 2242 | 1806 | 1441 | 1108 | 801 | 484 | 249 | 84 |
Product Surveillance Registry Results
Number of Leads Enrolled in Study | 2644 |
---|---|
Cumulative Months of Follow-Up | 113636 |
Number of Leads Active in Study | 974 |

US Acute Lead Observations
Cardiac Perforation | 21 |
---|---|
Conductor Fracture | 2 |
Failure to Capture | 24 |
Failure to Sense | 8 |
Impedance Out of Range | 18 |
Insulation Breach | 1 |
Lead Dislodgement | 54 |
Oversensing | 51 |
Unspecified Clinical Failure | 5 |

Product Characteristics
Lead Type | Defibrillation |
---|---|
Type and/or Fixation | Active Screw In |
Lead Function | Pacing/Sensing and Defibrillation |
Lead Placement | Transvenous |
Lead Tip Location | Right Ventricle |
Pace/Sense Polarity | True Bipolar/One Coil |
Serial Number Prefix | TAU |
---|---|
Lead Body Diameter | 8,2 FR |
Insulation | Silicone with Polyurethane Overlay |
Pace/Sense Connector Style | IS-1 BI |
High Voltage Connector | DF-1 |
Steroid | Yes |
Data as of October 12, 2018