6932 Sprint
Defibrillation
6932 Sprint
ALL
Information

Device Survival Probability

6932_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 138.0 mo                  
% Survival 99.2 98.3 98.3 98.3 97.7 97.7 97.7 96.7 96.7 95.5 92.4 92.4                  
#   364 305 242 203 157 125 105 91 81 68 56 54                  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 420
Cumulative Months of Follow-Up 25831
Number of Leads Active in Study 29
Information 11

Qualifying Complications

Extra Cardiac Stimulation 1
Failure to Capture 2
Failure to Sense 2
Impedance Out of Range 1
Oversensing 4
Other 1
Information

US Acute Lead Observations

Failure to Capture 2
Failure to Sense 2
Impedance Out of Range 1
Lead Dislodgement 4
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release1996-08-06
CE Approval Date
Registered USA Implants14900
Estimated Active USA Implants3402
US Market Release1996-08-06
CE Approval Date
Estimated WW Distribution22923
Information

USA Returned Product Analysis

Conductor Fracture 23
Insulation Breach 26
Crimp/Weld/Bond 0
Other 2
Information

WW Returned Product Analysis

Conductor Fracture 54
Insulation Breach 40
Crimp/Weld/Bond 0
Other 8
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Tines
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix TCA
Lead Body Diameter 7.8 FR
Insulation Silicone
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of November 2, 2015

www.medtronic.com/productperformance 376051