6931 Sprint Fidelis
Defibrillation
6931 Sprint Fidelis
ALL
Information

Device Survival Probability


6931_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr at 66 mo        
% Survival 98.2 96.2 93.0 89.2 83.8 81.6        
#   257 226 192 139 85 47        

System Longevity Study Results

Number of Leads Enrolled in Study 294
Cumulative Months of Follow-Up 13106
Number of Leads Active in Study 119
Information 37

Qualifying Complications

Conductor Fracture 19
Failure to Capture 3
Failure to Sense 1
Impedance Out of Range 7
Lead Dislodgement 2
Oversensing 5
Information

US Acute Lead Observations

Cardiac Perforation 1
Conductor Fracture 2
Failure to Capture 1
Failure to Sense 1
Lead Dislodgement 1
Oversensing 3
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release2004-09-02
CE Approval Date
Registered USA Implants8079
Estimated Active USA Implants3633
US Market Release2004-09-02
CE Approval Date
Estimated WW Distribution19802
Information

USA Returned Product Analysis

Conductor Fracture 504
Insulation Breach 1
Crimp/Weld/Bond 0
Other 8
Information

WW Returned Product Analysis

Conductor Fracture 1499
Insulation Breach 3
Crimp/Weld/Bond 0
Other 27
Information

Product Characteristics

Lead Type Defibrillation
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing and Defibrillation
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity True Bipolar
More
Serial Number Prefix LFL
Lead Body Diameter 6.6 FR
Insulation Silicone with Polyurethan Overlay
Pace/Sense Connector Style IS-1 BI
High Voltage Connector DF-1
Steroid Yes
Hide this content

Data as of March 12, 2013